Effect of Supply Voltage and Load Resistance on Verification Results of PTB220 Pressure Sensors
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Tests of supply voltages and load resistances on several PTB220 pressure sensors are carried out for the accuracy and reliability of verification results. The test results show that the measurement errors of pressure sensor analog output are increasing gradually with the increasing supply voltage in the range of 12 to 30 V. The maximum error variation is up to -028 hPa, while there is no effect for digital outputs. The analog measurement errors of the PTB220 pressure sensors are increasing with the increasing of load resistance at the low limit of pressure measurement range for load resistance variation between 10 kΩ and 500 kΩ. Based on this finding, suggestions for the modification of verification method of PTB220 pressure sensors are proposed accordingly.

    Reference
    Related
    Cited by
Get Citation

李建英,贺晓雷,崇伟,丁红英.电源及负载电阻对PTB220型气压传感器检定结果的影响[J].气象科技英文版,2015,43(2):181-185.